51
Technical Data
7.3 Environmental Conditions
7.3.1 Climate
7.3.2 Shock and Vibration
7.3.3 Mechanical Data
Climate
Operating temperature –10 … +50 °C (14 … 122 °F)
Storage and transportation –20 … +70 °C (–4 … 158 °F)
Max. altitude 2000m
Humidity 5 … 95% relative humidity, non-condensing
Tested acc. to IEC 60068-2-78, Cab, Damp
Heat: Temp. 40°C, duration 48h,
rel. humidity 95%
Dynamics
Vibration Tested according to IEC 60068-2-6;
frequency range 10 … 150 Hz; acceleration
2g continuous (20 m/s
2
); 20 cycles per axis
Shock Tested according to IEC 60068-2-27
(operating mode); 15g / 11ms, half-sinusoid,
3 shocks in each axis
Weight, Dimensions and Protection
Weight 2.6kg (5.7lbs) without accessories
Dimensions W x H x D 285 x 68 x 225mm
(11.2 x 2.7 x 8.9")
Degree of protection IP20 according to EN 60529