User Maintenance
8 User Maintenance
8.1 Care and Cleaning
The CT SB2 does not require any special maintenance or care. Clean the device
from time to time or as necessary using a cloth dampened with water or
isopropanol alcohol. Always disconnect the CT SB2 prior to cleaning!
8.2 Replacing Fuses
1. Turn off the CT SB2 and unplug the power cord.
2. Ground the test object, and disconnect it from the CT SB2. By disconnecting
it you prevent a possibly faulty test object from feeding power back into the
CT SB2.
3. Disconnect the CT SB2 from the CT Analyzer and, if applicable, from the
computer.
4. Locate the blown fuse on the side panel of the CT SB2 and replace it by an
identical fuse type: T2.0A H 250V (2.0 Amps slow-acting high breaking
capacity wire fuse 5 x 20mm). The CT SB2 has two fuses of the same type.