CT Analyzer User Manual
68
Remote interface
1
If your CT Analyzer is equipped with a USB interface and
a RS232 interface, use this option to display the Select
remote interface port page. On this page you can
select the interface used to connect the CT Analyzer to a
PC: USB or RS232.
If equipped with both interfaces, the CT Analyzer will only
communicate via the selected interface. It will not be
recognized by the computer if the CT Analyzer settings
do not match the interface used for connection.
Misc. Settings Use the Miscellaneous Settings page to
• enable or disable automatic test assessment if a
parameter marked with a star "*" in the CT-Object
card was not entered by the user prior to the test.
Parameters marked with a star are mandatory
parameters that are required for assessment (and
possibly for the calculation of further parameters).
• enable or disable continuous burden measurement.
• define the class multiplication factor.
• define the thresholds for the decision algorithms of the
guesser function.
Please refer to section 5.7.3 on page 70 for more
information.
1. Only available for CT Analyzer devices that are equipped with a USB interface (serial number
JHxxxx or newer).