Preventive Maintenance
SENSOR INSTRUCTIONS
Weekly cleaning of the sensor is recommended to maintain optimum performance. To maintain a
clean and clear appearance, spray a soft, lint-free, non-abrasive towel or cloth with a
commercially available household glass cleaner, such as Windex®, Glassex®, VISS®, and Mr.
Muscle® and gently clean the outside of the sensor glass.
DO NOT directly spray the sensor glass with glass cleaner; always spray a towel or cloth
with glass cleaner and then gently wipe the sensor glass.
DO NOT use an industrial-strength glass cleaner.
RUBBER ROLLER INSTRUCTIONS
The rubber roller may need cleaning periodically, as it is important to keep the roller free of
debris, adhesive buildup and other surface materials.
Locate the following supplies:
•
Dishwashing
liquid
• 1 cup of wate
r
• Lint-fre
e cloth
WARNING: Use only dishwashing liquid to clean the rubber roller. Do not clean
the roller with alcohol, window cleaner or petroleum-based products, as this
causes the rubber to crack.
Mix a few drops of dishwashing liquid into a cup of water and place a lint-free cloth into the
solution until the cloth is saturated. Ring out the cloth so that no water drips from the cloth. Wipe
down the rubber roller surface, removing any debris. As you wipe, turn the roller to clean the
entire surface. Then, rinse out the cloth and wipe down the rubber surface again to remove any
dishwashing liquid from the roller. Allow the roller to dry before using.
CALIBRATION CARD INSTRUCTIONS
Replace the Calibrated Conformance Standard Test Card every two years.
If you have any questions or concerns about the performance of the LVS-7500 External Sys
tem,
please call your local Omron Microscan distributor or Omron Microscan technical support: