LVS-7500 External System Operations Manual
LVS-7500 External System Operations Manual Page 99 of 129
Troubleshooting
Problem Description Possible Causes
System will not acquire images
The roll of labels is not moving.
Encoder is not in contact with moving roll of labels.
Encoder cable is not plugged in or is damaged.
The Readhead has lost power.
The Readhead to LVS-7500 interface cables are not
plugged in or are damaged.
System is acquiring every other label The label repeat is set too large.
Image has uneven lighting
An LED may have burned out in the sensor.
The sensor may have a label stuck to it
For Error Condition issues, refer to the “Step 6: Alarm Matrix” section and review the following topics:
Error Code Definitions
Common Causes for Error Conditions