GUIDANCE AND MANUFACTURERS DECLARATION
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP710N is intended for use in the electromagnetic environment specified below. The
customer or the
user of this OMRON BP710N should assure that it is used in such environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The OMRON BP710N uses RF energy only for its
internal fu
nction. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The OMRON BP710N is suitable for use in all
est
ablishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not Applicable.
Voltage
fluctuations/
flicker emissions
IEC61000-3-3
Not Applicable.