GUIDANCE AND MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BP710N is intended for use in the electromagnetic environment specified below. The
customer or the
user of this OMRON BP710N should assure that it is used in such environment.
Immunity test
IEC 60601
test leve
l
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood,
concre
te, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
Not Applicable. Not Applicable.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Not Applicable. Not Applicable.
Voltage dips, short
interruptions an
d
voltage variations on
power supply inputlines
IEC 61000-4-11
<5 % U
T (>95 % dip
in U
T)
for 0.5 cycle
Not Applicable. Not Applicable.
40 % U
T (60 % dip
in U
T)
for 5 cycles
70 % U
T (30 % dip
in U
T)
for 25 cycles
<5 % U
T (95 % dip
in U
T)
for 5 sec.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Note: UT is the A.C. mains voltage prior to application of the test level.