14
Cuff Selection and Connection
Warning
If a cuff is used on a patient with an infection, treat the cuff as medical
waste, or disinfect before reuse.
Caution
Do not use the cuff if it is damaged or has holes.
Use the appropriate cuff size to ensure correct measurements. If a
cuff that is too large is used, the measured blood pressure value tends
to be lower than the actual blood pressure. If a cuff that is too small is
used, the measured blood pressure value tends to be higher.
Note:
Make sure that the connectors are tightly connected.
Selecting the cuff
Measure the circumference of the patient’s arm and select the cuff
size that is appropriate for the circumference.
It is important to use the correct sized cuff for a patient in order to get
an accurate reading.
Select the cuff that is suitable for the patient from the cuffs below.
Cuff name
Arm circumference
FP LQFK
GS CUFF2 XL* 42 - 50 17 - 20
GS CUFF2 L* 32 - 42 13 - 17
GS CUFF2 M 22 - 32 9 - 13
GS CUFF2 S* 17 - 22 7 - 9
GS CUFF2 SS* 12 - 18 5 - 7
*Available as an optional accessory.
Connecting the cuff
Connect the cuff tube to the NIBP connector
on the unit.