Check before Use
Before turning on the power
Before turning on the power, check for the following
Ŷ ([WHUQDODSSHDUDQFH
The device or accessories are not deformed due to falling or other
impact.
The device is not dirty.
The device is not wet.
Ŷ $&DGDSWHU
The AC adapter is firmly connected to the connector on the device.
There are no heavy objects lying on the AC adapter cable.
The AC adapter cable is not damaged (core-wire exposure, breaks, etc.).
When turning on the power
When you turn on the power, check the
LCD display.
When the [START/STOP] or [Power
ON/OFF] button is pressed to turn on
the power, the right screen appears.
After turning on the power
After turning on the power, check for the following
Ŷ ([WHUQDODSSHDUDQFH
There is no smoke or odor coming from the device.
The device is not making any unusual noises.
Ŷ %XWWRQV
Press each button and check that it works.
Ŷ 1RQLQYDVLYHEORRGSUHVVXUH1,%3
Make sure that a suitable OMRON GS CUFF2 is attached (one that fits
the circumference of the patient’s arm).
The cuff tube is firmly connected.
The person checking the cuff should wrap the cuff around arm,
perform cuff measurement and check to see that blood pressure is in
the vicinity of normal measurements.
While measurement is in progress, intentionally activate “Body
movement detection function” to halt discharge by bending the relevant
arm and move body and during this halt check that cuff pressure does
not drop.