4
Caution
'RQRWLQVWDOOWKHXQLWLQWKHIROORZLQJORFDWLRQV
- Locations with dust, salt, or sulfur.
- Locations directly exposed to sunlight for extended periods of time (in
particular, do not leave in direct sunlight or near a source of ultraviolet
light for extended periods, as ultraviolet light will cause deterioration
of the LCD).
- Locations subject to vibration or shock.
- Near heaters.
Do not use in a location with noise-emitting equipment such as a room
with MRI, CT, X-ray or HF surgical equipment or an operating room.
Noise from the equipment may interfere with the operation of the unit.
Before use / during use
Warning
The unit complies with the EMC (Electromagnetic Compatibility)
standard (EN60601-1-2). As such, it can be used simultaneously with
multiple medical instruments. However, if instruments that generate
noise such as an electric scalpel or a microwave therapy device are
near the unit, check the operation of the unit during and after use of
these instruments.
If an error occurs or a measurement result is questionable, check the vital
signs of the patient by auscultation or palpation. Avoid relying solely on
the measurement results of the unit when judging the patient’s condition.
Only trained healthcare providers should use this device. Do not allow
patients to operate this device.
Properly connect the connectors and AC adapter cable.
Do not place objects or liquids on top of this unit.
&KHFNWKHIROORZLQJEHIRUHXVLQJWKHXQLW
- Make sure the AC adapter cable is not damaged (wires are not
exposed or broken), and the connections are firm.
For the AC adapter connected to the unit, supplies, and optional devices,
use only the standard accessories or OMRON-specified products. This
may damage and/or may be hazardous to the device.
Do not use in a location with moisture, or a location where water may
splash on the unit.
This unit is intended for use in physicians’ office, hospitals, clinics and
other medical facilities.
Do not use the unit if it emits smoke, an abnormal odor, or abnormal noise.
Do not bring cellular telephones or transceivers into the room where the
unit is installed or being used.
Do not connect multiple monitors to the same patient.
Do not connect the unit to a power outlet that is controlled by a wall switch.

Other manuals for Omron HBP-1120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron HBP-1120 and is the answer not in the manual?

Omron HBP-1120 Specifications

General IconGeneral
BrandOmron
ModelHBP-1120
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals