5.Handling Errors and Problems
5. Handling Errors and Problems
5.1
Error Messages
Error Display Cause Remedy
Cuff is under inflated.
Carefully read and repeat the
steps listed under section 3.4.
Movement during
measurement
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
Air plug disconnected.
Insert the air plug securely.
Refer to section 3.2.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.2.
Clothing is interfering
with the arm cuff.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.2.
Air is leaking from the
arm cuff.
Replace cuff with new one.
Refer to Chapter 7.
The arm cuff was
inflated above 299
mmHg when inflating
the cuff manually.
Release your finger from the
O/I START button before the
pressure reaches 299 mmHg,
when inflating the cuff
manually.
Refer to section 3.4.