EN2
Introduction
Thank you for purchasing the OMRON M7 Intelli IT (ALRU) Automatic 
Upper Arm Blood Pressure Monitor.
Operational principle of work
Oscillometric method:
The beat in the pulsation generated by the contraction of the heart is 
captured as the pressure inside the cuff to measure the blood 
pressure. If the cuff wrapped around the upper arm is pressurized 
sufficiently, the blood flow stops, but the beat of the pulsation is 
present and the pressure inside the cuff receives this and oscillates. 
Next, as the pressure inside the cuff gradually decreases, the 
oscillation of the pressure within the cuff gradually increases and 
reaches a peak. As the pressure within the cuff decreases further, the 
oscillation decreases from its peak. The pressure within the cuff and 
the relationship with the increase and decrease of the oscillation 
within the cuff in this series of processes are stored into memory, 
calculations are carried out, and the blood pressure value is 
determined. The oscillometric method does not determine the blood 
pressure value instantly like a microphone type automatic blood 
pressure gauge with the auscultation method, but rather determines it 
from the series of change curves as explained above. Therefore, it is 
not easily affected by external noise, an electric scalpel or other 
electro surgical instruments.
Safety Instructions
This instruction manual provides you with important information about 
the OMRON M7 Intelli IT (ALRU) Automatic Upper Arm Blood 
Pressure Monitor. To ensure the safe and proper use of this monitor, 
READ and UNDERSTAND all of these instructions. If you do not 
understand these instructions or have any questions, contact 
your OMRON retail outlet or distributor before attempting to use 
this monitor. For specific information about your own blood 
pressure, consult with your physician.
Intended Use
This device is a digital monitor intended for use in measuring blood 
pressure and pulse rate in adult patient population. This monitor 
detects irregular heartbeats during measurement and displays 
corresponding indication along with measurement results.