11. Guidance and Manufacturer’s Declaration
EN62
Table 2 - IMMUNITY TEST LEVELS
Phenomenon Basic EMC standard
IMMUNITY TEST
LEVELS
Electrostatic
discharge
IEC 61000-4-2 ±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
for enclosure port
Radiated RF
electromagnetic fields
IEC 61000-4-3 10 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
for enclosure port
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3 See table 3
Electrical fast
transients / bursts
IEC 61000-4-4 ±2 kV for Input a.c.
power port 100 kHz
repetition frequency
Surges Line-to-line IEC 61000-4-5 ±0.5 kV, ±1 kV for
Input a.c. power port
Conducted
disturbances induced
by RF fields
IEC 61000-4-6 3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM and
amateur radio bands
between 150 kHz and
80 MHz
80 % AM at 1 kHz
for Input a.c. power
port
Rated power
frequency magnetic
fields
IEC 61000-4-8 30 A/m 50 Hz and
60 Hz
for enclosure port
2895901-4A_HEM-7361T-ALRU_IM.fm Page 62 Wednesday, October 26, 2022 10:46 AM