F-5500 Thermal Mass Flow Meter Manual 06/17 - 2030 / 107023 Page 61
Model F-5500
OPERATION
F1 F2 F3 F4
WARNING! You must ensure that there is a no flow condition before proceeding. If you
are performing the test in a bottle, be sure to isolate the sensor in a closed container -
any air movement (even from a fan) may result in a false "fail" result.
Once process is stable, press YES (F3) key to begin the Zero CAL-CHECK
®
.
F1 F2 F3 F4
This test will take less than 5 minutes. The T=xx is a count down timer indicating how
much time is left to finish the test.
F1 F2 F3 F4
OR
F1 F2 F3 F4
OR
F1 F2 F3 F4
Upon test completion, the final value will be displayed along with the test result. The
test result may be:
• Pass < 0.80
• Warning > 0.80 < 1.0
• Fail > 1.0
If a "Warning" or "Fail" result is displayed, ONICON recommends that the probe be
removed from the pipe, the sensor cleaned, and the test be performed again. If the test
was performed in the pipe the first time, perform the test in a bottle for the re-test.
If a "Warning" or "Fail" result is displayed after performing the test a second time,
please call ONICON Service at 727-447-6140 for assistance.
Verifying ZRO
y.yyy T=xx
ZRO=2.321
Fail OK
Process Zero and
Stable? YES EXIT
Operation: Zero CAL-CHECK
®
Performing
Zero CAL-
CHECK
®
Test
ZRO=0.911
Warning OK
ZRO=0.259
Pass OK