Do you have a question about the Onicon F-5500 and is the answer not in the manual?
Category | Accessories |
---|---|
Model | F-5500 |
Manufacturer | ONICON |
Output | 4-20 mA, pulse |
Materials | Stainless steel |
Installation | Insertion |
Accessory Type | Flow Meter |
General safety precautions for installation, operation, and service.
Menu tree for performing the Zero CAL-CHECK® calibration validation test.
Explanation of how the thermal mass gas flow meter works.
Advantages of measuring mass flow over volumetric flow.
Description of ONICON's Direct Digitally Controlled sensor technology.
Ensuring actual pipe ID matches calibration certificate.
Checking pipe run requirements for insertion and inline meters.
Setting the correct probe depth for accurate readings.
Verifying correct power and 4-20mA wiring connections.
Confirming output signal wiring based on model type.
Precautions to observe during installation and operation.
Instructions for positioning the insertion flow meter for consistent flow profile.
Details on the process adapter fitting design for insertion and removal.
Step-by-step procedure for inserting the flow sensor and confirming ID.
Instructions for properly tightening compression fittings.
Safe procedures for removing the meter from the pipe.
Ensuring the flow direction indicator points correctly.
Ensuring sensor elements are aligned correctly in the process stream.
Instructions for positioning the inline flow meter for consistent flow profile.
Ensuring the flow indicator on the inline meter points in the direction of flow.
Critical safety warnings for wiring installations.
Specifications and recommendations for power wiring.
Details on power input requirements and recommendations.
Wiring diagram for 4-20mA output with customer-supplied power source.
Wiring diagram for 4-20mA output powered by the F-5500 unit.
Wiring for pulse/alarm output with customer-supplied power.
Wiring for pulse/alarm output powered by the F-5500 unit.
Wiring connections for Modbus or BACnet communication.
General information on HART connections.
Wiring for HART 4-20mA output with loop power from F-5500.
Wiring for HART 4-20mA output with customer-supplied power.
Setting 4mA and 20mA values and alarm fault levels.
Configuring alarm outputs for High/Low Flow and Temperature.
Settings for serial communication protocols (Modbus, BACnet, HART).
Accessing and programming the pulse/alarm output.
Setting flow cutoff, pipe diameter, filter, and alarms.
Setting the high flow rate alarm limit.
Setting the low flow rate alarm limit.
Steps to perform the Zero CAL-CHECK® test.
Results of the Zero CAL-CHECK® test and troubleshooting guidance.
Procedure to confirm NIST-traceable calibration at zero flow.
Mandatory safety precautions before performing maintenance.
Reasons for Zero CAL-CHECK® test failures and troubleshooting steps.
Further troubleshooting for installation issues like meter response and readings.
Information to diagnose and clear alarm codes with reasons and actions.