F-5500 Thermal Mass Flow Meter Manual 06/17 - 2030 / 107023 Page 65
Model F-5500
COMMUNICATIONS
Model F-5500
COMMUNICATIONS
F-5500 Commands Supported by Modbus
Bit Definition Comment
0 Power up indication Cleared when out of the power up sequence
1 Flow rate reached high limit threshold Set limit to zero to disable
2 Flow rate reached low limit threshold Set limit to zero to disable
3 Temperature reached high limit threshold Set limit to zero to disable
4 Temperature reached low limit threshold Set limit to zero to disable
5 Sensor reading is out of range Check sensor wiring
6 Gas mix error Gas mix must total 100%
7 Incorrect Settings Check settings
8 In simulation mode Set simulation value to 0 to disable
9 Pulse/alarm output is out of range Check pulse/alarm output settings
10 Analog 4-20 mA for flow/temp is out of range Check analog output settings
11 Not used Not used
12 Not used Not used
13 Not used Not used
14 CRC error Check parameters and reset CRC
15 Error in Total Reset total to clear alarm
Table 5.3: Status 2 Bits Definitions for Command 04, Modbus Address 30002
Bit Definition Comment
0 Pulse hardware detected
1 Busy Busy
2 Not used Not used
3 Not used Not used
4 Zero CAL-CHECK in process
5 Zero CAL-CHECK fail
6 Zero CAL-CHECK aborted
7 Zero CAL-CHECK warning
Table 5.2: Status Bits Definitions for Command 04, Modbus Address 30001
Preset Single
Register
Preset Single Register (Command 06)
This command is used to perform miscellaneous functions such as clearing the
totalizer and elapsed time. The register address is Modbus=40018 and the data to
write is described below.
Request:
<Meter Address> <Command code=06> <Register address high=0x00>
<Register address low=0x11> <Register data high=0x00> <Register data low
=0x02> <CRC high> <CRC low>