- 20 -
optek-Manual--1004-2006-02--TF16-US-2011-06-20
www.optek.com
Connection to converter C 4000
Note!
Detailed information on the connection of the sensor to a C4000-series
converter is given in the instruction manual of the converter. There you can as
well find the corresponding wiring plans.
Connection to the
converter
The following connections are on the back of the converter:
Fig. 14 Detector inputs and lamp outputs of the converter C4422
Letters stand for
• A - D Detector inputs
• E - F Connections for lamp outputs
The connections depend on the equipment of your converter and on how many
sensor you would like to connect.
Note!
Do not use other detector inputs and lamp outputs than given in table 9, even if
in some cases other combinations are possible. Thus, danger of mixing inputs
and outputs up is minimized.
20 2119
DETECTOR INPUTS
[C]
3
4
5
3
4
[D]
5
DETECTOR INPUTS
[A]
2
5
1
2
5
1
[B]
4 - 20 mA
OPTEK - DANULAT
REMOTE IN
R1
PA
M6
N
PE
L
R7
R8
RR
R4
R5
R6
R2
R3
115 / 230 V, 50 / 60 Hz, 50 VA
115 / 230 V, T 1,6 A
230V
!
24 V DC
IN 1, 2
acc. to
6
1031
MADE IN GERMANY
37
38
39
34
35
36
32
33
16
17
18
13
14
15
11
12
6
7
7
LAMP [F]
6
7
7
acc. to
cable
length
V DC
cable
length
0 / 4 - 20 mA
OUT 1, 2
24 V AC / DC
RELAY OUT
22
23
24
21
41
42
43
40
46
47
45
44
LAMP [E]
6
OUT 3, 4
V DC
Tab. 9 Connections
Number of
sensors sensor
type
Detector input connection of the converters
Lamp
output
Cable length max.
C4101 C4121 C4201 C4221 C4202 C4222 C4322 C4422
1 sensor
TF16
A / C
A / C A / C
A / C A / C
A / C E
250 m / 820 ft.
2 sensors
TF16
AF16
A / C
B
A / C
B
E
F
250 m / 820 ft.
2 sensors
TF16
AS16 / AS16-BT
A / C
B
A / C
B
E
F
250 m / 820 ft
50 m / 164 ft.
2 sensors
TF16
AS56
A / C
B
A / C
B
E
F
250 m / 820 ft
50 m / 164 ft.