CRYO-LINE Manual for installation and use
OPTIKON 2000
Code 131002 EN 2014-02-06 Rev.F
• Caution: N
2
O (Nitrogen Protoxide), if inhaled, is hazardous to the health: use the
system only if the gas is evacuated outside the operating room through a suitable
aspiration system.
• To ensure safe operation, do not install the system in a location where it may be
exposed to heating appliances or radiators, direct sunlight or any other source of
heat with extremely high temperatures.
• Install the CRYO-LINE system so that the ventilation openings are not obstructed.
• Do not dispose of the CRYO-LINE unit as non-separated municipal waste. Dispose of
it separately according to local electrical and electronic equipment disposal
laws/regulations.
• Do not use the system in the presence of flammable anaesthetic mixtures with air
or oxygen or Nitrogen Protoxide.
Electrical warnings
• The unit is already set up for use with a power supply voltage of 220V. Check that
the local voltage is the same as that predetermined on the unit. Otherwise, prepare
the unit for the local power supply voltage. In the countries where the local voltage
is 100 - 120 V, replace the fuses of the unit with the appropriate fuses supplied.
These operations must be carried out exclusively by appropriately trained
personnel.
• To reduce the risk of electric shocks, do not remove the protective cover. Request
the intervention of authorised service personnel.
• Connect the console of the CRYO-LINE unit to a mains having the characteristics
shown on the console’s rear panel. To guarantee safety, use a hospital-type power
cord and plug and connect it to a socket equipped with suitable earthing.
• Before connecting the unit to the mains, or disconnecting it, make sure that the
main switch of the console is off.
• When the CRYO-LINE unit is not in use, keep the main switch turned off.
• To allow cooling, the fuses must be replaced after the main switch has been turned
off for a few minutes.
• Replace the fuse with one of the same type, as indicated on the rear panel of the
console.