9.2 STERILIZATION
9.2.1 UNIT
1) The console, footswitch and cylinder cart cannot be sterilized.
9.2.2 CRYOSURGICAL PROBES
1) The cryosurgical probes can be sterilized in an autoclave. Refer to the specific user
instructions.
9.3 MAINTENANCE
9.3.1 UNIT
1) Store the CRYO-LINE cryosurgery system in a clean dry location at room temperature.
2) After use, disconnect the hose of the cylinder.
3) Periodically replace the cylinder filter.
4) To maintain optimal performance for the entire life of the equipment, an annual
maintenance and calibration program must be defined. Contact your Optikon 2000
authorised distributor / service centre.
NOTE
To keep the system safe, check the unit’s conformity with electric current leakage limits according to
standard EN60601-1 at least once a year. Contact the bioengineering division of your facility or your
Optikon 2000 authorised distributor / service centre.
9.3.2 ACCESSORIES
1) Avoid dropping the cryosurgical probes or handling them in an incorrect manner.
These accessories must be handled with the utmost care and carefully inspected
after each use to check for any damage or wear.
2) Periodically lubricate the external O-rings of the cryosurgical probe connectors, using
silicone grease or oil. DO NOT OVER-LUBRICATE.