10.2 MISCELLANEOUS PROBLEMS
The troubleshooting guide lists some of the anomalies that may occur in the unit’s
operation, the related symptoms and corrective actions. If the problems with the CRYO-
LINE surgical system persist even after having carried out the corrective actions below,
contact an OPTIKON 2000 authorised service centre.
SYMPTOMS CORRECTIVE ACTION
Mains power supply
CRYO-LINE completely inactive
a) The power cord is not connected. Connect the
cord to the console and the mains.
b) The fuse has blown. Replace it.
Gas leak
Leak that can be heard from the adapter
a) Adapter not tightened enough. Tighten it firmly.
b) Defective gasket. Replace the gasket.
Leak from the console
a) Pressure too high. Pressure which is too high
causes the opening of the safety valve inside the
instrument: connect to an appropriate pressure
source.
b) Defective valve. Call the Optikon technical
service.
a) Defective probe. Send the probe to Optikon for