CALIBRATION CONT.
26. The scale will count down from 10 to 0. Once 0 has been reached, the display will
show
CALEnd
.
27. Press
#
to continue. The display will now show
28. Press
ESC
to save and exit the setup menu.
29. The scale has now been calibrated. The display will show the value of the calibration
weight on the scale.
30. If the scale does not show the value of the calibration weight, check that the feet on
the platform are not screwed in too tightly, and verify that the platform is level.
31. Unload the scale; the display should read
000000
.
32. If the scale does not display 000000, check that the feet on the platform are not
screwed in too tightly, and verify that the platform is level.