digiBASE-E™ High-Performance Power-over-Ethernet Multichannel Analyzer / PMT Base for Scintillation Detectors
60
SHOW_ZERO_WIDTH
Reports the current width for the stabilizer zero peak. See also SET_ZERO_WIDTH,
SET_ZERO_CHANNEL, and SHOW_ZERO_CHANNEL.
Responses:
$C00001088<CR> Zero width has not been set.
$C00003089<CR> Zero width is 3 channels (lowest possible width).
... ...
$C00256100<CR> Zero width is 256 channels
START [seg-mask]
Starts data acquisition. The optional segment mask is provided for compatibility with other
MCBs and may be any value from 0 to 65535 but is ignored by the digiBASE-E.
Execution Warnings:
%000004073<CR> No segment selected (occurs with other warnings).
%000005074<CR> The acquisition is already started (no changes made).
%000006075<CR> A preset was exceeded (acquisition was not started).
%000032074<CR> High voltage is disabled.
The actual response record may be a combination of any of the above records depending on
the warning conditions.
STEP_OUTPUT
Causes a pulse to be output on the Change Sample output port. The level of the Change
Sample output is changed then returned to the level present when the STEP_OUTPUT
command was received. The duration of the intermediate level varies but is never less
than 2 μs. Use the SET_OUTPUT_HIGH or SET_OUTPUT_LOW commands to establish
an initial output level.
Responses:
$C00000087<CR> Initial and final level on Change Sample was low.
$C00001088<CR> Initial and final level on Change Sample was high.
STOP [seg-mask]
Stops the data acquisition. The optional segment mask is provided for compatibility with
other MCBs and may be any value from 0 to 65535 but is ignored by the digiBASE-E.
Execution Warnings:
%000005074 Acquisition already stopped (no changes made).