The eartips are disposable and therefore should not be cleaned or re-used.
Bone oscillator
Clean the bone oscillator between patients, e.g. with a non-alcohol based antibacterial wipe, such as Audiowipes.
Disposal
There are no special requirements for the disposal of eartips, i.e. they can be discarded according to local regulations.
15.3 Calibration
Annual calibration
The audiometer, headphones, bone oscillators, and sound field speakers must be calibrated once a year by your authorized
service department.
Caution Note that calibration has been performed only on the transducers supplied! If you wish to use any other
transducer for testing with the device, please contact your local distributor first.
16 Other references
For more information, see the online Help in OTOsuite, which contains detailed reference information about AURICAL
Aud and the OTOsuite modules.
For instructions on installing OTOsuite, see the OTOsuiteInstallation Manual, which you can find on the OTOsuite install-
ation medium (disk or memory stick).
17 Technical specifications
17.1 AURICAL Aud
Type identification
AURICAL Aud is type 1081 from GNOtometricsA/S.
Channels
Two separate and identical channels.
Frequency range
Insert earphones: Standard frequencies: 125 - 8000Hz
20 Otometrics - AURICAL Aud