EasyManua.ls Logo

Oval NPI45B20 - Pointer Movement is unusual.; Microswitches fail to produce any signal...; Potentiometer is faulty.... Pointer movement is all right.

Default Icon
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
 
①  Inspect for possible entrapment of air.
②  If trouble is attributable to the valve position
indicator, inspect in the same manner as in
“Valve position indicator is dead.”
 

①  Inspect associated equipment and field wiring
cables for condition.
②  Taking o󱐯 M3 screws (17), remove the indicator
cover (11) (Fig. 8.11). Disconnect one end of wire
leads from its terminal and make a continuity
(on and o󱐯) check relative to valve opening and
closure with multimeter (in the ohm range). If
faulty condition persists with no signal output,
inspect the microswitches.
③  Possible causes for the absence of signal output
are:
(A) An open in the internal wiring.
(B) Microswitch positions out of adjustment.
(C) Microswitches are defective. If this is the
case, replace with new ones.
[See paragraph (6) in Section 9.4 Assembly
Procedure on page 19 for adjustment.]
 
①  Inspect related equipment and eld wiring cables for condition.
②  Possible causes for trouble are:
(A) An open or poor connections in the internal wiring.
(B) An open in the resistor. If this is the case, replace with a new one.
[See paragraph (5) in Section 9.4 Assembly Procedure on page 19 for adjustment.]
 
①  It is recommended that the valve position indicator be disassembled and inspected on a regular basis
(once every three years or so).
②  Seals, including O-rings, should be replaced with new ones each time a periodic disassembly and
inspection is performed.

Related product manuals