7. Elektromagnetilise ühilduvuse vastavustabel
7.1 Kaabli pikkus
Cables & accessories Length max. Type of test In accordance with
Cables / Cords <3 m
Emission RF CISPR 11, classe B
Harmonic currents emission IEC 61000-3-2
Voltage uctuation and ickers IEC 61000-3-3
Electrostatic discharge immunity IEC 61000-4-2
Radiated immunity
Electromagnetic elds
IEC 61000-4-3
Immunity to interference against fast
transient electrical disturbance/burst
IEC 61000-4-4
Immunity to fast transients IEC 61000-4-5
Conducted immunity
RF conduct interference
IEC 61000-4-6
Radiated immunity
Magnetic elds
IEC 61000-4-8
Immunity to volrage drops, short inter-
ruptions and voltage variations
IEC 61000-4-11
7.2 Soovituslik miinimumvahekaugus
Meditsiiniseade on mõeldud kasutamiseks elektromagnetilises keskkonnas, kus on raadiosageduskiirgusest
tingitud häired kontrolli all.
Meditsiiniseadme kasutaja või paigaldaja saab aidata elektromagnetilisi häiringuid vältida, järgides minimaalset
vahekaugust, olenevalt raadiosagedusedastusseadmete maksimaalsest võimsusest. Kaasaskantavaid
raadiosageduse sideseadmeid (sh välisseadmeid, nagu antennikaableid ja väliantenne) ei tohi kasutada lähemal
kui 30 cm (12 tolli) kaugusel seadme osast, sealhulgas tootja määratud kaablitest. Vastasel juhul võivad nende
seadmete jõudlused muutuda.
7.3 Elektromagnetkiirgus
Emission test Standard Electromagnetic environment
Electromagnetic radiation distur-
bances.
(Radiated emission)
(CISPR 11)
Groupe 1 The Nocospray 2 uses RF energy only
for its internal function.
Terminal supply disturbance voltage
(Conducted emissions)
(CISPR 11)
Classe B The Nocospray 2 is suitable for use in
all establishments, including domes-
tic establishments and those directly
connected to the public low voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonics emissions
(IEC61000-3-2)
N/A
Voltage / Flicker test
(IEC 61000-3-3)
N/A