7. EMC COMPLIANCE
7.1 Cable length
Cables & accessories Length max. Type of test In accordance with
Cables / Cords <3 m
Emission RF CISPR 11, classe B
Harmonic currents emission IEC 61000-3-2
Voltage uctuation and ickers IEC 61000-3-3
Electrostatic discharge immunity IEC 61000-4-2
Radiated immunity
Electromagnetic elds
IEC 61000-4-3
Immunity to interference against fast
transient electrical disturbance/burst
IEC 61000-4-4
Immunity to fast transients IEC 61000-4-5
Conducted immunity
RF conduct interference
IEC 61000-4-6
Radiated immunity
Magnetic elds
IEC 61000-4-8
Immunity to volrage drops, short inter-
ruptions and voltage variations
IEC 61000-4-11
7.2 Recommended separaon distance
The medical device is intended for use in an electromagnetic environment in which disturbances due to RF
radiation are controlled.
The user or installer of the medical device can help to avoid elecromagnetic interference by maintaining a
minimum distance, depending on the maximum power of the radio frequency transmission equipment. RF
portable communications equipment (including peripherals such as antenna cables and external antennas)
should not be used closer than 30cm (12 inches) to any part of the UNIT TEST, including the cables specied by
the manufacturer. Otherwise, the performance of these devices may be impaired.
7.3 Electromagnec emissions
Emission test Standard Electromagnetic environment
Electromagnetic radiation distur-
bances.
(Radiated emission)
(CISPR 11)
Groupe 1 The Nocospray 2 uses RF energy only
for its internal function.
Terminal supply disturbance voltage
(Conducted emissions)
(CISPR 11)
Classe B The Nocospray 2 is suitable for use in
all establishments, including domes-
tic establishments and those directly
connected to the public low voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonics emissions
(IEC61000-3-2)
N/A
Voltage / Flicker test
(IEC 61000-3-3)
N/A