7.5 Elektromagnesk immunitet, RF
Immunity test IEC60601 test level Compliance level Electromagnetic environ-
ment
RF portable communications equipment (including peripherals such as antenna cables and external antennas) should
not be used closer than 30cm (12 inches) to any part of the UNIT TEST, including the cables specied by the manufactu-
rer. Otherwise, the performance of these devices may be impaired.
Radiated RF
(IEC61000-4-3)
3V/m
80MHz to 2,7 GHz
80% MA at 1kHz
10v/m
80MHz to 2,7 GHz
80% MA at 1 kHz
3V/m
80MHz to 2,7 GHz
80% MA at 1kHz
10v/m
80MHz to 2,7 GHz
80% MA at 1 kHz
Home health environment
and a
professional health esta-
blishment
Proximity elds issued by
RF wireless communication
devices
(IEC 61000-4-3 provisory
method)
9v/m
710MHz, 745MHz, 780MHz,
5240MHz, 5550MHz,
5785MHz
27V/m
385MHz
28V/m
450MHz, 810MHz, 870MHz,
930MHz, 1720MHz,
1845MHz, 1970MHz,
2450MHz
9v/m
710MHz, 745MHz, 780MHz,
5240MHz, 5550MHz,
5785MHz
27V/m
385MHz
28V/m
450MHz, 810MHz, 870MHz,
930MHz, 1720MHz,
1845MHz, 1970MHz,
2450MHz
Home health environment
and a
professional health esta-
blishment
Conducted RF 3V
150KHz to 80MHz
6V in ISM band and band
between 0.15 MHz and 80
MHz including amateur
radio band.
80% MA at 1KHz
3V
150KHz to 80MHz
6V in ISM band and band
between 0.15 MHz and 80
MHz including amateur
radio band.
80% MA at 1KHz