ET
7.4 Kindlus elektromagnet- ja magnetkiirguse suhtes
Immunity test IEC60601 test level Compliance level Electromagnetic environ-
ment
Electrostatic discharges
(IEC61000-4-2)
± 8Kv contact
± 15Kv air
± 8Kv contact
± 15Kv air
Home health environment
and a
professional health esta-
blishment
Electrical fast Transients
IEC61000-4-4)
± 2Kv for power supply lines ± 2Kv for power supply lines
± 1Kv for input/output lines
Home health environment
and a
professional health esta-
blishment
Surge
(IEC61000-4-5)
± 1Kv line(s) to line(s)
± 2Kv line(s) to earth
± 1Kv line(s) to line(s)
± 2Kv line(s) to earth
Home health environment
and a
professional health esta-
blishment
Magnetic eld at power
frequency
(IEC-61000-4-8)
30A/m 30A/m Home health environment
and a
professional health esta-
blishment
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
(IEC61000-4-11)
0% UT for 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°.
0% UT for 1 cycle
70% UT
For 25 cycles at 50Hz
For 30 cycles at 60Hz
Monophasé à 0°
0% UT for 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°.
0% UT for 1 cycle
70% UT
For 25 cycles at 50Hz
For 30 cycles at 60Hz
Monophasé à 0°
Home health environment
and a
professional health esta-
blishment
Voltage interruptions
(IEC61000-4-11)
0% UT
For 250 cycles at 50Hz
For 300 cycles at 60Hz
0% UT
For 250 cycles at 50Hz
For 300 cycles at 60Hz
Home health environment
and a
professional health esta-
blishment