EasyManua.ls Logo

Oxypharm NOCOSPRAY - Chemical Substance Safety; Handling Chemical Substance Precautions

Oxypharm NOCOSPRAY
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
EN
The device diuses a chemical substance. Due to this, the following recommendations must imperatively be
respected:
Users are required to read the accompanying documents.
Decontamination process need to be monitored.
The device must be used solely indoors.
Do not inhale the product.
Do not smoke in proximity to the device
Do not direct the device towards an incandescent area.
Do not use in the presence of an inammable anaesthetic mixture.
Decontamination process need to be monitored
Evacuate the room during diusion.
Respect the contact times provided in the usage protocols (minimum 30 minutes) before returning into the treated
area.
In case of emergency, if a person must return into a treated area before treatment is over, they must be equipped
with Personal Protection Equipment: hearing protection and PPE in accordance with the product’s safety data
sheet.
Before allowing safe entry into the treated area for those not protected with PPE, a H202 gas detection unit can be
used to measure any residual gas.
Entry must only be permitted when the ppm count falls below that of the requisite levels stipulated by local laws of
exposure times in the workplace and mentioned in the product’s SDS.

Table of Contents