There must not be any strange noises when the bed is
traveling.
There must not be any strange vibrations when the bed is
traveling.
The board stopper must be locked properly.
There must not be any deformation, damage, or detachment
of bumpers, resin parts or elastic parts.
There must not be any peeling of labels, and the text must be
legible.
Back raise operation (p.17 to
p.23)
Proper operation can be achieved.
There must not be any strange noises.
There must not be any scratches and deformation at the
mattress base, frame, or any other parts.
Hi-lo and TR/RTR operations
(p.17 to p.22)
Proper operation can be achieved.
There must not be any strange noises.
Knee raise operation (p.17 to
p.23)
Proper operation can be achieved.
There must not be any strange noises.
Board mounting condition
(p.14)
There must not be any remarkable rattling.
Bed side-rail operation (p.28 to
p.29)
Proper operation can be achieved.
There must not be any strange noises or remarkable rattling.
The operation must not be unusually difficult.
The operation must not be unusually difficult.
The bed must not move when the casters are locked.
When in steering position, swiveling of the steering caster
must be locked.
Caster locking operation (p.26
to p.27)
The bed must move smoothly when the lock is released.
Proper operation can be achieved.
Visual check of electrical parts
Cable must not get tangled in the bed.
Plugs and cables must have no scratches or damage.
There must not be any scratches and damage on the
handheld controller.
For safety in use, inspect the product regularly, at least once a year. The product exhibits wear and
deterioration according to the frequency and conditions of use. Inspect for loose parts, the operation of
moving parts on the bed, and the presence of damage regularly. When an irregularity has been noticed
or when a detailed inspection is desired, contact Paramount Bed or your distributor.
REGULAR INSPECTION AND DAILY MAINTENANCE