●Back raise cannot move in
up direction even when the
handheld controller, the
side-rail integrated panel or
the patient control panel
buttons are pressed.
●Check if emergency CPR function
is operated previously.
●Press the [↓] button of the back
raise adjustment buttons on the
handheld controller or the side-rail
integrated panel for more than 5
seconds in power on status.
●Noises when emergency
CPR function is operated.
●Check if the CPR lever fully
pull/grip when in use.
●Pull/Grip fully to the CPR lever
when in use.
■Product ID label locations
Product ID labels are affixed to the foot end of the main frame and on the inside of the footboard.
The lifetime of this equipment is typically eight (8) years. Lifetime is defined as the period during
which the product will maintain the specified performance and safety, provided it has been maintained
and operated in conditions of normal use in accordance with the requirements in this instruction manual.
When requesting repairs or contacting Paramount Bed or your distributor, please provide
the number on the product ID label.