●Abnormal ticking sound
occurred during back raise
operation.
●Check if emergency CPR function
is operated previously.
●Press the [↓] button of the back
raise adjustment buttons on the
handheld controller or the side-rail
integrated panel for more than 5
seconds in power on status.
●The bed does not move
even when the handheld
controller buttons are
pressed.
●Check if the bed was operated
continuously.
●Reset the protect condition.
●The bed does not move
even when the side-rail
integrated panel and the
patient control panel
buttons are pressed.
●Check if the bed was operated
continuously.
●Reset the protect condition.
●Check if the Lock-out switch is
in the lock position.
●Rotate the switch to the unlock
position.
●The bed cannot be
moved.
●Check if the power plug is
plugged into the outlet.
●Plug the power plug into the
outlet.
●The bed does not move
even when the handheld
controller, the side-rail
integrated panel or the
patient control panel
buttons are pressed.
●The power plug is plugged into
the outlet but the bed still does
not move.
●Unplug the power plug from
the outlet and plug it in again.
●Plug another electrical
device into the outlet to
check that power is available.
●Check if the casters are locked.
●Check if the side-rail integrated
panel and the patient control
panel lock function is OFF.
●Press the lock button to switch
the side-rail integrated panel and
the patient control panel lock
function OFF.
●The bed cannot be
positioned correctly.
●Check if there is nothing in the
moving parts on the bed.
●Beeping alert sound occurred
when the handheld controller,
the side-rail integrated panel
or the patient control panel
buttons are pressed.
●Check if the battery is charged.
●Plug the power plug into the outlet.
●The board will not come
off.
●Check if the board is locked.
●Unlock the board stoppers.
●The bed does not move in
the intended direction
when the patient is being
transported.
●Check if the caster is in the
steering position.
●Unlock the caster in the
steering position.
Problems may not always indicate faults. Check the following once again before requesting repair.
■If the problem persists after taking corrective actions, stop using the bed immediately, disconnect the
power plug from the outlet, and contact Paramount Bed or your distributor for repair.
■Bed damaged in earthquakes, fires, floods or other incidents must be inspected and repaired by
Paramount Bed or your distributor. Otherwise, there is a risk of electric shock or fire due to
short circuit or current leakage from electrical parts, of injury due to malfunctioning caused by
deformation of the bed.