Note: UT is the AC power voltage before applying the test level.
250/300 cycles for test level 0% UT
0.5 cycle for test level 0% UT
(Phase: at 0°, 45°, 90°, 135°, 180°,
225°, 270°and 315°)
1 cycle for test level 0% UT
and
25/30 cycles for test level 70% UT
(Single phase: at 0°)
250/300 cycles for test level 0% UT
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
Voltage dips, instantaneous power
failure, and voltage fluctuation in
power supply cable
IEC 61000-4-11
1 cycle for test level 0% UT
and
25/30 cycles for test level 70% UT
(Single phase: at 0°)
0.5 cycle for test level 0% UT
(Phase: at 0°, 45°, 90°, 135°, 180°,
225°, 270°and 315°)
Short-term power failure
IEC 61000-4-11
Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The A5 Series is intended for use in the following electromagnetic environment. The customer or user of the A5
Series must use it in an environment specified here.
Electrostatic discharge (ESD)
IEC 61000-4-2
Contact discharge: ±8kV
Air discharge: ±2kV,±4kV,
±8kV, ±15kV
Contact discharge: ±8kV
Air discharge: ±2kV,±4kV,
±8kV, ±15kV
Line to line: ±0.5kV, ±1.0kV
Line to earth: ±0.5kV, ±1.0kV, ±
2.0kV
Power supply line:±2kV
Input/output Lines:±1kV
Electrical fast transient/ burst
IEC 61000-4-4
Power supply line:±2kV
Input/output Lines:±1kV
Line to line: ±0.5kV, ±1.0kV
Line to earth: ±0.5kV, ±1.0kV, ±
2.0kV