Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, must be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The A5 Series is intended for use in the following electromagnetic environment. The customer or user of the A5
Series must use it in an environment specified here.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic Environment - Guidance
Radioactive RF
electromagnetic field
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
80 amplitude
modulation (1 kHz)
3 V/m
80 MHz to 2.7 GHz
80 amplitude
modulation (1 kHz)
RF Conductivity
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V
ISM bandwidth
between
0.15 MHz and 80
MHz
80 amplitude
modulation (1 kHz)
3 V
0.15 MHz to 80 MHz
6 V
ISM bandwidth
between
0.15 MHz and 80
MHz
80 amplitude
modulation (1 kHz)
58
a. Field strengths from fixed transmitters, such as those of base stations for radio (cellular/cordless) telephones,
land mobile radios, amateur radios, AM and FM radio broadcasts, and TV broadcasts cannot be predicted
theoretically with accuracy.To determine the electromagnetic environment due to fixed RF transmitters, an
electromagnetic survey of the site should be considered. If the measured field strength in the location in which
A5 series is used exceeds the applicable RF compliance level above, A5 series should be observed to verify
whether or not it is operating appropriately. If an abnormal operation is seen, take additional measures as
needed, such as changing the location of use of the A5series.
b. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection fromstructures, objects, and people.