Pulse
modulation
18Hz
1.8
0.3
28V/m
450MHz
28V/m
810MHz
28V/m
800 to
960
GSM 800/900,
TETRA 800,
iDen 820,
CDMA 850,
LTE band 5
Pulse
modulation
18Hz
2
0.3
TETRA 400
27V/m
385MHz
27V/m
380 to
390
0.3
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The A5 series is intended to be used in an electromagnetic environment where RF radio communication devices
are used. The electromagnetic interference between the RF communication device and the A5 series can be
prevented by maintaining the minimum distance between the two as recommended below in accordance with the
maximum output and frequency of the communication device.
Distance
(m)
Maximum
output
(W)
Modulation
Immunity test
IEC 60601-
1-2
Test level
Compliance
level
Bandwidth
(MHz)
Service
9V/m
710MHz
9V/m
745MHz
9V/m
780MHz
9V/m
704 to
787
LTE band 13, 17
Pulse
modulation
217Hz
0.2
0.3
28V/m
430 to
470
GMRS 460,
FRS460
Pulse
modulation
18Hz
2
28V/m
870MHz
28V/m
930MHz
28V/m
1720MHz
28V/m
1700 to
1990
28V/m
2450MHz
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE band 1, 3, 4,
25; UMTS
Pulse
modulation
217Hz
2
0.3
28V/m
1845MHz
28V/m
1970MHz
0.2
0.3
9V/m
5500MHz
9V/m
5785MHz
Proximate
electromagnetic
field
IEC 61000-4-3
9V/m
5240MHz
9V/m
5100 to
5800
WLAN 802.11
a/n
Pulse
modulation
217Hz
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE band 7
Pulse
modulation
217Hz
2
0.3
2400 to
2570
28V/m
59
SPECIFICATIONS
14

Table of Contents