Bulletin TI-H2PEMPD-D Installation, Operation and Maintenance Manual
Hydrogen Gas Generator H2PEMPD Series
www.parker.com/fns/balstonlabgenerators
2
1-800-343-4048
Parker Balston
Technical Information Bulletin
Please note that all Parker gas generators have a warranty period
of 1 year from the date of shipment
WARNING!
The H2PEMPD-1300 hydrogen generator, operating at full
ow, will consume the entire water reservoir volume in
approximately 60 hours.
If the user expects to run the generator, unattended, for greater than 60
hours then the auto water feed feature MUST BE used. See user’s guide for
details.
Failure to use auto water feed will result in loss of ow and pressure, and
could result in damage to generator.