Bulletin TI-H2PEMPD-D Installation, Operation and Maintenance Manual
Hydrogen Gas Generator H2 PEMPD series
24
1-800-343-4048 www.parker.com/balston
Generator/Cell HIGH
pressure
Contact tech support/
PSU. PCB, Transducer
fault
Reset/replace, check
connection
No output pressure signal
Reset/replace, check
connection
Reset/ check sensor
connection
Cell separator water level
HIGH
Check/replace sesnor or
connection
Calibration file out of date
Calibrate conductivity
sensor
Reset/contact tech
support
Pump fault – no cell water
circulation
Check/replace pump or
connection
Power supply fan not
rotating
Case cooling fan not
rotating
Reset/replace the
external fan
Reservoir level sensor
faulty
Reset/check connector,
replace sensor
Purifier temp sensor faulty
Reset/check connector.
Replace heater
O2 FLOAT CONNECT
ACT 100%