Bulletin TI-H2PEMPD-D Installation, Operation and Maintenance Manual
Hydrogen Gas Generator H2 PEMPD series
25
1-800-343-4048 www.parker.com/balston
Purifier temp rise rate =
low
Off
Reset/check connector.
Replace heater
Purifier temp HIGH
Off
Reset/check connector.
Replace heater
Purifier temp LOW
Off
Reset/check connector.
Replace heater
Service time
On
Perform service. Use
applicable kit
O2 vent is blocked
Off(System lock)
Reset/remove blockage &
reset
JP3 on options board
shorted
Off
Remove short/reset
TEMP RATE
ACT 100%
OVER TEMP
ACT 100%
LOW TEMP
ACT 100%
SERVICE REQUIRED
ACT 100%
O2 VENT BLOCK
ACT 100%
REMOTE STOP
ACT 100%