52
5.1
Setup, Continued:
SUMMARY: The Setup Menu consists of the following items; the following sections provide further
explanation:
SCALE: This is used to set the proper communications for the various scale manufacturers. While
Mettler is preferred, other scales from Ohaus or Sartorius may be used. Submenu allows setting of
Scale Manuf., Units, Alarm, and Tare. Default = “METLER”.
CLOCK: Set the time of day (military), day, month, and year. Print Enable allows choice of Time of
Day, or Relative (Run) Time for printout and display. In most cases the clock will be set at the factory
for the destination time zone. Default = Time of Day.
SYSTEM TEST: Allows testing of the I / O’s of the LabTec™, requires purchase of IQ / OQ Document.
Use Test Mode if needed for trouble shooting.
TEST MODE: Allows independent testing of the LabTec™ I / O’s. Motor, Keypad, Scale, Pressure,
Temperature, Valves, Analog, TTL switches.
ETHERNET: Allows setting of the IP Address, Subnet Mask and Gateway values for Modbus TCP / IP
communication via the Ethernet Port. Modbus communication not available at the time of this printing.
PRINTER: Select communications parameters for Parker printer (P/N 080-095) or
PC. Default
settings are Print Time (Default = 30 sec.), Type (Seiko), Baud Rate (9600), Stop Bits (2),
Parity (None),
Word Length (8), Print Delay (0 sec). Controls these settings for both the Printer and
USB ports.
TEMPERATURE: Allows setting of a Temperature Alarm and Limit, as well as an Offset for the
SciTemp
®
Temperature sensor if needed.
PRESS. SENSOR: Allows user to Zero all three SciPres
®
pressure sensors. (Span is used for factory
calibration.) Source is used to select the desired sensor for related alarms and control in all modes.
Units: Choose from psi (default), bar, or kpa. Range: Default is 60, can be set lower, will require
re-calibration of the input. Contact Parker Customer Service for assistance.
PUMP: Select the following user preferences: Keypad Beep: (On / Off), Switch Configuration:
(Level / Pulse), Switch Polarity: (Normal / Inverted), TTL1:On-Off: (Yes / No), Motor Start:
(Hard / Soft-Ramp), Motor RPM: (3400, 600, 160, 8), Pump Head: ( P e r i s t a l t i c / R H 1 / R H 0 / R H 0 0 ) ,
Pump Tubing: (13 / 14 / 16 / 25 / 17 / 18 / 15 / 24 / 35), Power Up: (Mode / Menu / Run), External Run-
Stop: (Pulse / Level), ASCII Feedback (On / Off), Factory Reset: (Resets all variable parameters to
their original factory defaults).
SCALE2 and SCALE3: Not utilized with the LabTec™. Must remain set to None”. Do not remove the
covers on the unused scale ports.