5.1 Setup: Scale
UP
- S E T U P - S C A L E
UP DOWN
SELECT
SELECT
- SCALE - ALARM
UP DOWN
SELECT
SELECT
DOWN
TARE WEIGHT: OFF
UP DOWN SELECT
SCALE MANUF: Select the appropriate value for the scale in use. Options: Ohaus, Ohaus2, Ohaus3,
Mettler, Metler2, Sartor, Sartor2. Default = Mettler. Proper configuration of the scale’s parameters is
required, as well as correct interface cable.
•
Ohaus3: Adventurer Pro, Adventurer, (New version available in 2015)
•
Ohaus2: Adventurer, old (pre-2002), Explorer, Explorer Pro.
•
Mettler: Viper, Series 4, IND560.
•
Metler2: Speedweigh, Panther.
•
Sartor2: Current default Sartorius setting, all series.
ALARM: Triggered if communication with the scale is lost. Options: Pump Stop, Alarm Only or Off.
Default = Pump Stop.
TARE WEIGHT: Determines if the system tares the scale upon pressing Execute and Run in the main
operational modes. This is counterintuitive. Options: On, Off. Default = OFF, which causes the system
to tare the scale. ON will cause the tare to not occur.
UNITS: Select from: .001 gm, .01 gm, 0.1 gm, Kg, T, Lbs, OzT, Oz, C, Dwt. Default = .01 gm.
A Mettler Toledo ICS425 series scale is the most frequently purchased scale sent with the system. If a
different scale is required, please contact Parker for configuration information.
- SCALE - SCALE MANF.
UP DOWN SELECT
SELECT
DOWN
SCALE MANF: OHAUS3
UP
DOWN SELECT
- SCALE - UNITS
UP DOWN
SELECT
SELECT
UNITS: .01GM
UP DOWN SELECT
- SCALE - TARE WEIGHT
UP DOWN SELECT
SELECT
DOWN