5.11 Mettler Toledo ICS425 Scale Parameters
To enter the Setup mode, press and hold the “Transfer” key until the word CODE appears on the
display. Then press “Zero”, “Zero”, “Zero”, “Transfer”, and an interactive menu screen is
presented. Use the displayed arrows and the related buttons to navigate the menu and confirm
the following settings are enabled.
Level 1
Level 3
Level 3
Selected Menu Item
SCALE
Disp Unit & Res Disp Unit 1
g
Disp Unit 1
g
Disp Resolution OFF (sets to default value)
Zero Filter
AZM
OFF
Vibration Medium
Process
Dosing
Stability Standard
Comm
COM1 Mode Dialog
Param RS 232
Baud
9600
Parity 8 None
H. Shake OFF
When exiting the Menu a prompt to save occurs, press OK to save the changes.
Setup Mode, select “METLER”, see SETUP: SCALE: MANUF. By making this selection, the system will
implement the correct parameters for communicating with this scale.
You will also need P/N 080-072ICS Cable, Mettler Balance, ICS M-12.
Scales purchased with the system through Parker will be configured and tested to-
gether with the
LabTec™ as a system at the factory.
Note: The bold parameters listed are the only custom settings needed for communication with
Parker systems. All others should remain at factory defaults. Consult your scale
manual for help
navigating through the scale menu. If stability is an issue, adjust the Vibration
and Stability
settings as needed.