EasyManua.ls Logo

Penlon Prima SP2 - Device Classification and Labelling

Penlon Prima SP2
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Power Outlets
Auxiliary power outlets Four outlets:
5 A per outlet
10 A maximum total
Fuses T5AH ceramic (¼ x 1¼ inch) high breaking capacity fuse on
Live and Neutral on each outlet
Overload Protection 10 A thermal circuit breaker incorporated within the power
outlet On/Off switch
Electromagnetic Compatibility
The Prima SP2 meets the requirements of EN 60601-1-2
(electromagnetic compatibility - requirements and tests)
4.11 Device Classification and Labelling
Type B Applied Part
Degree of protection against electric shock
This symbol denotes: Type B equipment
Class 1 Classification
Type of protection against electric shock: Class 1
IPX0 Ingress Protection
Classification according to the degree
of protection against ingress of water: IPX0 (not protected)
Labelling
This symbol denotes: Refer to the User Manual
18
SPECIFICATION
دﯾﯾﺎﻣرﻓ ﮫﻌﺟارﻣ.WWW.ELM-FAN.IRتﯾﺎﺳ ﮫﺑ رﺗﺷﯾﺑ دراوﻣ دوﻠﻧاد یارﺑ
To download more items, go to ELM-FAN.IR