6.2 A200SP Absorber
6.2.1 Removing the Absorber
Assembly
CAUTION
The canister (1) may contain condensate.
Before removing the absorber from the pole-
mount assembly, refer to the A200SP user
manual for instructions on draining the canister
and disposing of absorbent.
1. Set the brakes on the anaesthetic
machine front castors.
2. Refer to the A200SP user manual
and disconnect all cable connectors
and hoses, including the mains lead
from the heater unit (2) (if fitted).
3. Slacken the knob (3), and lift the
absorber assembly from the pole-
mount.
USER MAINTENANCE
32
1
3
2