5.9 Ventilator
Always follow the pre-use check procedures
given in the ventilator instruction manual.
Check all hose and tubing connections for
gas tightness.
Check all wiring connections for correct
fitment and security.
AV-S Ventilator Interface
Check the Interface Cable
Check for correct fitment of the interface
cable at the socket (1) on the rear panel of
the Prima SP2, and the connection on the
AV-S control unit, marked ‘Prima master’.
Check the interface system function
a) Turn the Gas Delivery Switch ON.
The ventilator will power-up.
b) While the Prima SP2 power is ON, the
ventilator can be turned off and on, using the
ventilator On/Off switch.
c) Turn the Gas Delivery Switch to OFF.
The ventilator will power-down.
1
29
PRE-USE CHECKS