b. Using Cells
Show the cell positions which are selected for measurement. Remove the cells by and
switch positions by and keys.
c. Multi Zero
If checked, zero will be measured all the selected positions. If not, zero will be only measured at
the 1
st
cell position among the using cells.
NOTE:
To prevent inadequate baseline (zero) measurement because of the difference in transmittance among
cuvettes or any other reasons, Multi Baseline (Multi Zero) measurement is recommended.
d. Temperature (°C): Enter the returning temperature after the experiment is finished.
e. Monitor
Select temperature that used to monitor during measurement. Options include: Block, Probe1 or
Probe2.
f. Start (°C): Enter the start temperature for the measurement.
g. End (°C): Enter the end temperature for the measurement.
h. Interval (°C)
Enter the measurement interval temperature. For instance, if you enter 5°C, the sample will be
measured from start temperature to end temperature every 5°C.
i. Rate (°C/min): Enter the heating (or cooling) temperature rate in each temperature range.
NOTE:
The rate has to be set under or equal to interval value.
j. Hold (min)
Enter the holding time. For example, if you set 1 min, measurement will start one
minute passed after temperature reaches at the set point.
k. Error Range(°C/min)
It shows the temperature tolerance between the sampling and monitored temperature. The
measurement will start when the temperature tolerance reaches within the set Error Range.
Example: Set the starting temperature at 20°C and error range with 1°C then, the
measurement will start when the temperature reaches between 19°C and 21°C.
▶Temperature Display
l. Select which temperature will be displayed on the panel:Block, Probe 1 or Probe 2.
The selected temperature is only displayed in real time.
m. Offset:
This function is used to adjust temperature. It is only used for manufacturing process, so
do not modify the values.