09931286E
23
10. After completing the parameter setup, select Apply and OK. Then the LED for ON-LINE is turned
on, and it will start heating up or cooling down to the Start temperature in the Multicell Peltier
tab.
11. Insert blank solutions into the both reference and sample cell holder. Select Baseline and then
baseline will be measured when it reaches the set Start Temperature.
NOTE:
To monitor the probe temperature, the probes should be immersed in the sample, or to use the block
temperature for monitoring, the cell lid should be closed tightly.
a. Start Temp (°C): The starting temperature of experiment.
b. SP Temp (°C): The set temperature to go to the next measurement temp.
c. Block Temp (°C): The temperature of Cell Block.
d. Probe 1 Temp (°C): The temperature of Probe 1.
e. Probe 2 Temp (°C): The temperature of Probe 2.
f. Monitor: Selected temperature is being monitored.