g. Error Range (°C): It shows the temperature tolerance between the sampling and monitored
temperature. The measurement will start when the temperature tolerance reaches within the set
Error Range.
h. Hold Time: It shows the set holding time.
i. Elapsed time (s): It shows the elapsed time to reach the sampling temperature.
j. Status: It shows the status of experimental progress.
12. Remove blank solution from sample cell holder and insert sample solution into the sample cell
holder. Select Sample, then the measurement will be started after the temperature reaches the
target temperature.
NOTE:
To pull out the cell easily, use the cell lifting knob.
Figure 23 Pull out the cell
13. After the measurement is finished, the results are displayed in the result window. Save or print the
results as required.
Other Measurement Modes
NOTE:
Peltier Temperature Control Unit is available in all methods, however, unlike Thermal Denaturation, the
measurement is only performed with isothermal state
.
1. Install the 6-Position Peliter Controlled Cell Changer referring to the chapter Installation.
2. Close the sample compartment cover and turn on the instrument.
NOTE:
When executing System Self Test, the 6-Position Peltier Controlled Cell Changer has to be installed.
Otherwise, the instrument can be damaged electrically and does not operate properly.
3. Double-click on the UV Express folder and select one of the modes except Thermal Denaturation.
In this case, open kinetics mode as example.
4. Select OK after finishing the System Self Test.
5. Select New to open a new window and select OK.