50 . High-Performance Lambda Spectrometers Hardware Guide
3. Undo the two retaining screws located under the blanking caps on the top of the
detector (Figure 18), which are accessible using the long, posidrive screwdriver supplied
with the instrument.
Figure 18 Undoing the retaining screws
4. Remove the detector compartment cover by carefully lifting vertically using the two
hand holds, as shown in Figure 19.
Figure 19 Removing the detector compartment cover
5. Move the detector unit carefully to the right and remove it from the spectrometer.
The detector unit is installed by performing this procedure in reverse, tightening the
retaining screws and making sure the blanking caps are in place.
WARNING
Make sure you screw down the detector unit as this creates the earth
bonding for electrical safety.
Details of the installation procedure for a sphere, the Three Detector Module or the Universal
Reflectance Accessory (URA) are given in the relevant manual on the
High-Performance Lambda Spectrometers User Manuals
CD
(L6050009).