EN – 23
EN Start-up
Purge and indicator
connections
Serie Item Description Connection
C2
C1
1
Gas ballast purge
1/8’ NPT
2
Oil casing purge
1/8’ NPT
3
Oil pressure sensor
connection *
M 10 X 1
4
Purge - Bubbler
1/8’ NPT
5
Oil casing temperature
sensor connection *
M 12 X 1
Specific equipment: HP valve
(C2 Series)
The C2 pumps are equipped with HP and LP elastomer valves (FPM™ according to
standard NFT 40-002). In certain applications with fluorinated gases, HP valve may
harden and become brittle and compromise pump performances. Alternatively, the
pump may be equipped with a plastic, corrosion-resistant HP valve on request. On
request we can supply these specific components depending on your needs.
Equipment HP Valve material Tightness Part Number
Standard Elastomer Installation tight
053443
On request
Plastic * Installation not tight
065057
Plastic + special o–ring Installation tight
065160
* If a plastic valve is used, the pump’s anti-suckback operation cannot be
guaranteed and an anti-suckback device must be installed in the pumping
line to ensure tightness when the pump is not operating.
* Sensors are customer supplied.