Operating instructions HE lite
Page 14 of 28 Revision 9 © 2017 Pfeuffer GmbH
7.2 Offset
Each single product can be calibrated by
means of an additive constant factor
(Offset) within a possible range of -2 % to
+2 % (in steps of 0.1 %). In addition to this
the instrument offers the possibility to
enter a factor, allowing linearly ascending
corrections.
Select the function »Offset« and then
press the central key »SET«.
and move to the next figure by pressing
Set figure »4« by means of the left arrow
key
:
and move to the next figure by pressing
Enter the figure »3« by means of the left
arrow key
:
7.3 Code entering
A code inquiry prevents unintentional
adjustments of the instrument.
»CODE 001« is indicated on the display.
Enter the figure »2« by means of the left
arrow key
:
Now the correct code »342« appears on
the display and must be confirmed by
pressing
Battery
Offset
Average
Set
Code: 002
Set
Code: 042
Set
Code: 342
Set